How to configure the probe of the sterilizer
How to configure the probe of the sterilizer, where is it located?
There are at least 2 probes in the OSR sterilizer, of which 1 fixed probe is installed at the lowest point as a control probe; there is at least 1 mobile probe in the cabinet, which can be placed arbitrarily. The number of mobile probes is increased according to user requirements or model size. The standard configuration of OSR is 1 mobile probe for cabinets with a length of less than 2 meters, and 2 mobile probes for cabinets with a length of more than 2 meters. The mobile probe is placed arbitrarily according to the needs, but in actual use, it is generally fixed at the bottom of the equipment or hung on the inner wall. Prevent the probe from being damaged when the sterilization cart is pushed in and out. (The water bath sterilizer has 2 fixed probes, one at the discharge port and one at the water inlet, corresponding to the inlet and outlet of the plate heat exchanger respectively.)