This series of water bath leak detection sterilizer uses high temperature circulating water as the sterilization medium, and is mainly used for water spray sterilization and leak detection cleaning of ampoules, vials and other injections and oral liquid products. Purified water is circulated through a water pump and heated by a plate heat exchanger to perform water spray sterilization on items. The equipment has good temperature uniformity and a wide temperature control range (80-126℃), while avoiding secondary contamination during the working process. The leak detection method combining vacuum and colored water ensures a 100% scrap detection rate.

Category:

This series of water bath leak detection sterilizer uses high temperature circulating water as the sterilization medium, and is mainly used for water spray sterilization and leak detection cleaning of ampoules, vials and other injections and oral liquid products. Purified water is circulated through a water pump and heated by a plate heat exchanger to perform water spray sterilization on items. The equipment has good temperature uniformity and a wide temperature control range (80-126℃), while avoiding secondary contamination during the working process. The leak detection method combining vacuum and colored water ensures a 100% scrap detection rate.

Common engineering parameter

Steam pressure0.4~0.6MpaDesign pressure0.17~0.25Mpa
Water pressure0.2~0.4MpaWorking pressure0.15~0.23Mpa
Compressed air pressure0.5~0.7MpaWorking temperature126℃
Preset programGlass bottle / Injection / Oral liquid / Soft bag / Plastic bottle sterilization program and etc.

CORE VALUES

  • Innovative design
  • High-ended finishing
  • Reliability of all processes
  • Safe and secure devices
  • Easy to validate
  • Customer oriented solutions
  • Use of innovative materials & components
  • Energy savings
  • User friendly solutions

Material standard

PartMaterial PartMaterial
ChamberSS304Chamber supportSS304
StiffenerQ245RPipelinesSS304
Door base-plateSS304Internal trolleySS304
RackSS304External trolleySS304
Door sealing gasketSilicon rubberOuter casing housingSS304

Parameter & Size

ModelInner chamber size (L×W×H)mmOverall size (L×W×H)mm Double doorNet weight kgSteam consumption kg/cycleControl power AC220/50Hz/kwSupply power 3-AC380/50Hz/kwWater consumption kg/cycle
OSR-CKS-1.21500×680×11801840×2100×200021001200.55.5730
OSR-CKS-1.51870×680×11802210×2100×200023001500.57910
OSR-CKS-2.01700×1000×12002040×2600×200027001900.58.51220
OSR-CKS-2.52100×1000×12002440×2600×200030002400.59.51220
OSR-CKS-3.02500×1000×12002840×2600×200036002900.511.51440
OSR-CKS-4.03340×1000×12003680×2600×200042003800.5131920
OSR-CKS-5.03340×1000×15003680×2600×220049004800.515.52400
OSR-CKS-6.04000×1000×15004340×2600×220057005800.5182880